Description
- Flue gas analysis for all liquid, gaseous and solid fuels
- Latest ECOx sensor technology with 5 years guarantee for O2 and CO cells
- Features Bluetooth® and QR code data transmission to the EuroSoft® connect App
- Extendable with CAPBs® units and high-quality accessories for all types of service measuring tasks
- All-in-one: Flue gas analysis, (differential) pressure measurement, (differential) temperature measurement and air flow measurement
The MULTILYZER® STx NOx Set perform accurate measurements of O2, CO/H2, NO, NO2, temperatures, pressure (draft) as well as precise calculation of various parameters such as lambda, combustion efficiency (Eta), flue gas loss (qA), dew point, NOx and CO2. The real NOx value is measured with the NO and NO2 sensor combination.
Scope of delivery | |
M04743210 | MULTILYZER® STx (O2, CO, NO, NO2, P-D7, BLE) |
511010 | Option: Datenlogger |
510921 | Basic handle with draft, 2.4 m |
523229 | Interchangeable Probe 300 mm |
523493 | Power supply unit |
523506 | Charging cable |
500141.1 | Air sensor 80 mm |
500208 | Filter set |
511035 | microSD card 4GB with card reader |
523503 | Case |
O2 (oxygen) | |
Measuring range | 0 to 21.0 vol.-% |
Accuracy | ±0.2 vol.-% RDG |
Resolution | 0.1 vol.-% |
Response delay (t90) | < 30 seconds |
CO (carbon monoxide) with H2-compensation | |
Measuring range | 0 to 10,000 ppm |
Accuracy | ±5 ppm (< 50 ppm) |
±5 % RDG (> 50 ppm) | |
Resolution | 1 ppm |
Response delay (t90) | < 60 seconds |
CO2 (carbon dioxide) calculated | |
Measuring range | 0 to CO2 max. (depending on fuel) |
Accuracy | ±0.2 vol.-% |
Resolution | 0.1 vol.-% |
Response delay (t90) | < 30 seconds |
Flue gas temperature | |
Measuring range | 0 to 1,150 °C |
Accuracy | ±1 °C + 1 digit (< 300 °C) |
±1 % RDG (> 300 °C) | |
Resolution | 0.1 °C |
Response delay (t90) | < 30 seconds |
Ambient air temperature | |
Measuring range | -20 to 200 °C |
Accuracy | ±3 °C + 1 digit (-20 to 0 °C) |
±1 °C + 1 digit (0.1 to 200 °C) | |
Resolution | 0.1 °C |
Response delay (t90) | < 30 seconds |
Draft/differential pressure | |
Measuring range | ±70 hPa (draft) |
±150 hPa (diff. pressure) | |
Accuracy | ±0.02 hPa (< 2.00 hPa) |
±1 % RDG (> 2.00 hPa) | |
Resolution | 0.01 hPa (< 20 hPa) |
0.1 hPa (> 20 hPa) | |
Max. overload | 1350 hPa |
Response delay (t90) | < 10 seconds |
Pitot | |
Measuring range | 0.5 to 70 m/s |
Accuracy | ±0.8 m/s |
Resolution | 0.1 m/s |
Response delay (t90) | < 20 seconds |
Barometric pressure | |
Measuring range | 500 to 1.150 hPa |
Accuracy | ±10 hPa |
Resolution | 1 hPa |
Response delay (t90) | < 60 seconds |
Option: CO2 (carbon dioxide) | |
Measuring range | 0 to 20 vol.-% |
Accuracy | ±1 vol.-% |
Resolution | 0.1 vol.-% |
Response delay (t90) | < 60 seconds |
NO (nitrogen oxide) | |
Measuring range | 0 to 5,000 ppm |
Accuracy | ±5 ppm (< 50 ppm) |
±5 % RDG ( > 50 ppm) | |
Resolution | 1 ppm |
Response delay (t90) | < 60 seconds |
NO2 (nitrogen dioxide) | |
Measuring range | 0 to 500 ppm |
Accuracy | ±10 ppm (< 50 ppm) |
±10 % RDG (> 50 ppm) | |
± 5 ppm (< 100 ppm)1) | |
± 5 % RDG (> 100 ppm)1) | |
Resolution | 0.1 ppm |
Response delay (t90) | < 60 seconds / < 180 seconds1) |
Option: SO2 (sulfur dioxide) | |
Measuring range | 0 to 5,000 ppm |
Accuracy | ±10 ppm (< 200 ppm) |
±5 % RDG (> 200 ppm) | |
±5 ppm (< 200 ppm)1) | |
±5 % RDG (> 200 ppm)1) | |
Resolution | 1 ppm |
Response delay (t90) | < 60 seconds / < 180 seconds1) |
Option: COhigh (carbon monoxide) | |
Measuring range | 0 to 4.0 vol.-% (= 40,000 ppm) |
Accuracy | ±10 ppm |
±5 % RDG | |
Resolution | 0.001 vol.-% |
Response delay (t90) | < 60 seconds |
Option: NOlow (nitrogen oxide) | |
Measuring range | 0 to 300 ppm |
Accuracy | ±2 ppm (< 40 ppm) |
±10 % RDG (> 40 ppm) | |
Resolution | 0.1 ppm |
Response delay (t90) | < 60 seconds |
Option: SO2x (sulfur dioxide) | |
Measuring range | 0 to 2,000 ppm |
Accuracy | ±5 ppm (< 200 ppm) |
±10 % RDG (> 200 ppm) | |
Resolution | 0.1 ppm |
Response delay (t90) | < 60 seconds / < 180 seconds1) |
1) with extended flue gas treatment (e.g. MAXISYSTEM ST)
Reviews